YR05496 color spectrophotometer is developed by Kalstein with independent intellectual property rights.It uses 45/0 geometrical optical structure, comply with CIE No.15, GB/T 3978. With human body engineering mechanics and humanized operation design, YR05496 color spectrophotometer has stable performance precise measurement and powerful functions in the leading position of the same industry.
With 400~700nm, YR05496 handheld color spectrophotometer features with high accuracy DeltaE*ab 0.04, and 8mm aperture makes measurement so easy. It can save 1000 Standards, 15000 Samples and you also can input your own LAB or XYZ value into the instrument. With SQC8 software, you can do better color control management. Plus, it is widely used in plastic, electronic, paint, ink, textile, garment, printing and dyeing, food, medical, cosmetic, industries, scientific research institutes, schools and laboratories.
Model |
YR05496 |
Illumination/observation system |
45/0 method (45 ring-shaped illumination, vertical viewing) Comply with CIE No.15,GB/T 3978. |
Integrating sphere Size |
Φ58mm |
Light Source |
combined LED sources |
Sensor |
Silicon photodiodearray |
Wavelength range |
400~700nm |
Wavelength pitch |
10nm |
Reflectance range |
0~100% |
Measuring Aperture |
Φ8mm |
Color Space |
CIE LAB,XYZ,Yxy,LCh,CIE LUV |
Color difference Formula |
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00 v |
Other Chromaticity Data |
Whiteness – WI(ASTM E313,CIE/ISO,AATCC,Hunter), Yellowness-YI(ASTM D1925,ASTM 313), TI(ASTM E313,CIE/ISO), Metamerism Index (Mt) Colour Stain, Color Fastness, Color Strength, Opacity |
Observer |
2°/10° |
Illuminant |
D65,A,C,D50,D55,D75,F2,F6,F7,F8,F10,F11,F12 |
Display Data |
Spectral Value/Graph, Colorimetric Value, Color Difference Value/Graph, PASS/FAIL Result, Color Offset, Color Simulation |
Measurement Time |
1.5s |
Repeatability |
Spectral Reflectance: standard deviation within 0.1%(400~700nm: within 0.2%) Colorimetric Value: Standard deviation within DeltaE*ab 0.04 (Measurement conditions: white calibration plate measured 30 times at 5 seconds intervals after white calibration was performed.) |
Inter Instrument Agreement |
Within DeltaE*ab 0.2 (Average for 12 BCRA Series II color tiles) |
Dimension |
L*W*H=90*77*230mm |
Weight |
600g |
Battery |
Li-ion battery. 5000 times within 8 hours. |
Lamp Life |
5 years, more than 1.6 million measurements |
Display Screen |
TFT 3.5inch, Capacitive Touch Screen |
Interface |
USB/RS-232 |
Data Memory |
1000 Standards, 10000 Samples |
Operating Temperature |
0~40℃(32~104°F) |
Storage temperature |
-20~50℃(-4~122°F) |
Standard Accessory |
Power Adapter, Li-ion Battery, Operating Instruction, CD-ROM (containing management software), Data Line, White and Black Calibration Cavity, Protective Cover and Wrist Strap |
Optional Accessory |
Micro Printer, Powder Test Box |
Notes |
The specifications are subject to change without notice. |
Weight | 3 kg |
---|---|