Pump type |
Drip Type Infusion Pump |
Pumping Form |
peristaltic fingers |
Adjustable Fall Rate Range |
1 ~ 400 drops / min (step: 1 drop / min) |
Adjustable Volume Rate Range |
1 ~ 1200 ml / h (when 1 ~ 99.9 ml / h, step: 0.1 ml / h; greater than 99.9 ml / h, step: 1 ml / h) |
Flow Accuracy |
Within 3% or so (using manufacturer’s designated IV kit or high quality calibrated IV kit) |
Mechanical precision |
within 2% or so |
Purge Rate |
800ml / h |
KVO rate |
1 ml / hour (1 ~ 300 ml / hour) |
3 ml / h (more than 300 ml / h) |
Infusion Volume Range |
1 ml ~ 9999.9 ml (step: 0.1 ml) |
Maximum Accumulated Volume |
9999.9 ml (step: 0.1 ml) |
Time interval |
1 min ~ 9999 min (step: 1 min) |
Occlusion Alarm Threshold |
Height: 800 mmHg + -200 mmHg (106.7 kPa + -26.7 kPa) |
Medium: 500 mmHg + -100 mmHg (66.7 kPa + -13.3 kPa) |
Low: 300 mmHg + -100 mmHg (40.7 kPa + -13.3 kPa) |
Alarm |
Infusion Full, Empty, Bad Signal, Malfunction, Occlusion, Door Open, Air Bubble, Low Battery, Setup Error, AC Power Off, Idle. |
Air Bubble Detector Method |
ultrasonic wave, sensitivity: greater than or equal |
25 l |
Fuse |
F1AL / 250V (“F” indicates fast, “L” indicates low breaking power), 2 pieces (installed inside the pump). |
Power supply |
AC 85 ~ 265V, 50 / 60Hz; |
Internal Battery |
11.1V rechargeable lithium-ion battery. Capacity: greater than or equal to 2000 mAh; |
The pump can run more than 4 hours at a flow rate of 25ml / h after charging for 8 hours (average flow rate specified by GB 9706.27-2005). |
Energy |
less than or equal to 40VA |
Size |
188mm (length) x 198mm (width) x 228mm (height) |
Weight |
2.2 kg |
Environmental requirements |
Transport |
Ambient temperature: -30 ° C ~ + 55 ° C |
Relative humidity: 20% ~ 95% (non-condensing) |
Atmospheric pressure: 70kPa ~ 106kPa |
Storage |
Ambient temperature: -30 ° C ~ + 55 ° C |
Relative humidity: 20% ~ 95% |
Atmospheric pressure: 70kPa ~ 106kPa |
Operation |
Ambient temperature: + 5 ~ + 40 ° C |
Relative humidity: 20% ~ 90% |
Atmospheric pressure: 86kPa ~ 106kPas |
Calibration function |
disposable IV set of any brand is suitable for this pump |
Three Working Modes |
Rate / volume / time mode |
Alarm |
Infusion Full, Empty, Bad Signal, Malfunction, Occlusion, Door Open, Air Bubble, Low Battery, Setup Error, AC Power Off, Idle. |