Description
The YR01871 nucleic acid extractor is an automated nucleic acid extraction system that incorporates our innovative rotary mixing technology (RMT), which can significantly reduce the aerosol generated during the purification process, minimize the risk of false positive results caused by cross-contamination, and increase the accuracy of experiment results. The YR01871 nucleic acid extractor is compatible with deep 96-well plates and six-tube strips, making this nucleic acid extraction system capable of delivering high throughput extraction and reducing waste without the conventional drawbacks caused by fixed throughput. Flexible and efficient, the YR01871 nucleic acid extractor offers 1 to 96 samples per cycle.
7-inch color touch screen:
Built-in 7 inch full color LCD display, easy to operate the experiment.
High performance and efficient extraction:
The YR01871 nucleic acid extraction system is a 6*16 extraction module with a special deep 96-well plate and 6-strip tubing designed to ensure high throughput and reduce reagent waste. The YR01871 allows extraction of 1 to 96 samples per cycle.
Innovative rotary mixing technology:
Based on Tianlong’s innovative rotary mixing technology (RMT), the YR01871 can reduce the aerosol generated during the experiment and minimize the risk of false positives caused by cross-contamination, while being super quiet during operation.
Negative pressure system with HEPA filtration:
With a negative pressure ventilation design, the YR01871 automated nucleic acid extraction system is available for replaceable HEPA filtration to ensure that the extracted air is free of biohazards.
High purification and reliable results:
A residual amount of magnetic beads of less than 1% increases the confidence in the result of your experiment.
Technical specifications
Model | YR001871 |
Throughput | ene-96 |
Reaction Volume | 30-1000uL |
Sample Volume | 200uL |
Compatible Consumables | 96-deep-well plates |
customized 6 strip tube | |
Inter-well Extraction Difference | CV ≤3% |
Rotational Speed | ≤3000rpm |
Heating Temperature | Lysis heating: room temperature to 120°C |
Elution heating: room temperature to 120°C | |
Mixing Method | Rotary mixing |
Operation Mode | 7-inch full-color LCD touch screen operation |
Program Storage | Up to 1000 programs can be storaged |
Protocol Management | Flexible to create, edit and delete protocols |
Automatic Control | Motor-driven automatic opening and closing of the experiment cabinet |
Magnetic Bead Residue | ≤1% |
Power Failure Protection | Choose freely whether or not to continue the experiment when the power is on again after cutting off |
Contamination Control | Negative pressure HEPA exhaust filter module; |
Built-in UV disinfection module | |
Connection Port Type | USB port |
Weight | 45kg (net) |
Instrument Dimensions | 510mm*490mm*480mm (W*L*H) |
Power Supply and Power Consumption | AC 100V-240V, 50/60HZ; 600VA |